User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: Simpson 260-6 DATE: 16-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 29 NUMBER OF LINES: 104 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A CONFIGURATION: Fluke 5220A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Lay the Simpson on its back for this calibration. 1.002 DISP Zero the Meter Movement using the Center Screw. 1.003 ASK- R F 1.004 HEAD {Volts DC and Linearity Tests} 1.005 DISP Connect the Red Test Lead to the 250mV input terminal 1.005 DISP of the Simpson. Connect the Black Test Lead to the 1.005 DISP COMMON - Terminal of the Simpson. 1.005 DISP 1.005 DISP Connect the other end of the Test Leads to the 1.005 DISP 'Binding Posts to Double Banana Plug Adapter' 1.005 DISP Red to Red and Black to Black. 1.006 DISP Set the Function Selector switch on the left hand side 1.006 DISP of the Test Instrument to + D.C. 1.007 DISP Set the Test Instrument Range to the 2.5V Range. 1.008 5100 250 250mV 2/ 2W 2.001 DISP Move the Red Test Lead from the 250mV input terminal 2.001 DISP to the +1V input terminal above the function switch. 2.002 5100 1 1V 2/ 2W 3.001 DISP Move the RED Test Lead from the +1V input terminal to 3.001 DISP the + terminal near the bottom left hand corner. 3.002 5100 2.5 2.5V 2/ 2W 4.001 DISP Set the Test Instrument Function switch to - D.C. 4.002 5100 2.5 -2.5V 2/ 2W 5.001 DISP Set the Test Instrument Function switch to + D.C. 5.002 ASK+ R 5.003 5100 10 1V 2/ 2W 6.001 5100 10 5V 2/ 2W 7.001 5100 10 10V 2/ 2W 8.001 5100 50 50V 2/ 2W 9.001 5100 250 250V 2/ 2W 10.001 5100 500 500V 2/ 2W 11.001 DISP Move the RED test lead to the 1000V input terminal 11.001 DISP at the lower right of the Simpson. 11.002 5205 1000 1000V 2/ 2W 12.001 HEAD {Volts AC Tests} 12.002 DISP Set the Test Instrument Function switch to A.C. 12.003 ASK- R 12.004 5205 1000 1000V 3/ 400H 2W 13.001 ASK+ R 13.002 DISP Move the RED test lead from the 1000V input terminal 13.002 DISP to the + terminal near the bottom left corner of 13.002 DISP the meter. 13.003 5100 500 500V 3/ 400H 2W 14.001 5100 250 250V 3/ 400H 2W 15.001 5100 50 50V 3/ 400H 2W 16.001 5100 10 10V 3/ 400H 2W 17.001 5100 2.5 2.5V 3/ 400H 2W 18.001 HEAD {Frequency Response Tests} 18.002 5100 2.5 2.5V 3/ 200H 2W 19.001 5100 2.5 2.5V 3/ 60H 2W 20.001 HEAD {Current DC Tests} 20.002 DISP Set Test Instrument Function switch to + D.C. 20.003 5100 1 1mA 2/ 2W 21.001 5100 10 10mA 2/ 2W 22.001 5100 100 100mA 2/ 2W 23.001 5100 500 500mA 2/ 2W 24.001 ASK- R 24.002 DISP Set the Test Instrument Range to 50V (u amps) 24.003 DISP Move the Red wire from the Test Instrument + Input 24.003 DISP to the 50 uAMPS Input. 24.004 5100 50 50uA 1.5/ 2W 25.001 DISP Set the Test Instrument Range to 10MA (amps) 25.002 DISP Move the Red wire from the meters 50uAMPS 25.002 DISP Input to the +10A Input and the Black lead to the 25.002 DISP -10A Input of the meter. 25.003 5220 10 10A 2/ 2W 26.001 HEAD {Resistance Tests} 26.002 DISP Move the Red wire from the Test Instruments +10A Input 26.002 DISP to the + Input near the bottom left hand corner. 26.002 DISP Move the Black wire from the -10A Input to the 26.002 DISP - COMMON Input Terminal of the Simpson. 26.003 DISP Set the Range to {Rx1}. 26.004 DISP Short the wires from the Test Instrument together at 26.004 DISP the Test Interface Panel. 26.004 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 26.004 DISP Remove the short. 26.005 RESF 10Z +12.5% -11% 2W 27.001 DISP Set the Range to {Rx100}. 27.002 DISP Short the wires from the Test Instrument together at 27.002 DISP the Test Interface Panel. 27.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 27.002 DISP Remove the short. 27.003 RESF 1000Z +8% -7.5% 2W 28.001 DISP Set the Range to {Rx10,000}. 28.002 DISP Short the wires from the Test Instrument together at 28.002 DISP the Test Interface Panel. 28.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 28.002 DISP Remove the short. 28.003 RESF 100000Z +8% -7.5% 2W 29.001 END